콘텐츠로 건너뛰기
Merck

HPLC Analysis of Caffeine and Taurine on Ascentis® Express HILIC Column using Evaporative Light Scattering Detection (ELSD)

HPLC Analysis of Caffeine and Taurine on Ascentis® Express HILIC Column using Evaporative Light Scattering Detection (ELSD) application for HPLC

Materials

분석 컬럼

제품 번호
설명
가격

Ascentis® Express HILIC, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 5 cm × 3 mm

표준 물질

제품 번호
설명
가격

Taurine

≥99%

Caffeine

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

CONDITIONS

column

Ascentis Express HILlC, 5 cm x 3.0 mm I.D., 2.7 μm particles (53967-U)

mobile phase

[A] 100 mM ammonium formate, pH 3.0 with formic acid; [B] acetonitrile; (10:90, A:B)

flow rate

0.6 mL/min

column temp.

25 °C

detector

ELSD, 55 °C, 3.5 bar nitrogen

injection

2 μL

sample

1000 μg/mL in 10:90, water: acetonitrile

설명

분석 메모

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of caffeine and taurine. Elution is by aqueous normal phase (ANP) as an increase in organic will increase retention.

법적 정보