GC Analysis of Tricyclic Antidepressants in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber

GC Analysis of Tricyclic Antidepressants in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materials

SPME 파이버

제품 번호
설명
가격

SPME fiber assembly Polydimethylsiloxane (PDMS)

df 100 μm(PDMS, needle size 24 ga, for use with manual holder

분석 컬럼

제품 번호
설명
가격

Equity-1 Capillary GC Column

L × I.D. 30 m × 0.32 mm, df 0.25 μm

CONDITIONS

sample/matrix

1 mL urine (1 μg each analyte + 50 μL 5 M NaOH) in 7.5 mL vial

SPME fiber

polydimethylsiloxane, 100 μm (57300-U)

extraction

headspace, 15 min, 100 °C (sample incubated 30 min)

desorption process

3 min, 280 °C

column

100% methyl polysiloxane, 30 m × 0.32 mm I.D., 0.25 μm (Supelco equivalent, Equity-1, 28055-U)

oven

100 °C to 300 °C at 20 °C/min

inj. temp.

splitless (splitter opened after 1 min), 280 °C

detector

FID, 280 °C

carrier gas

helium, 30 cm/sec