GC Analysis of Volatiles in Milk (Before and After Exposure to Sunlight) on Supel-Q PLOT after SPME using a 75 μm PDMS/Carboxen Fiber

GC Analysis of Volatiles in Milk (Before and After Exposure to Sunlight) on Supel-Q PLOT after SPME using a 75 μm PDMS/Carboxen Fiber suitable for GC, application for SPME

Materials

SPME 파이버

제품 번호
설명
가격

SPME fiber assembly Carboxen/Polydimethylsiloxane (CAR/PDMS)

df 75 μm(CAR/PDMS, for use with manual holder, needle size 24 ga

관련 제품

제품 번호
설명
가격

SPME Fiber Holder

for use with manual sampling

SPME Fiber Holder

for use with CTC CombiPAL and Gerstel MPS2

분석 컬럼

제품 번호
설명
가격

Supel-Q PLOT Capillary GC Column

L × I.D. 30 m × 0.32 mm, average thickness 15 μm

표준

제품 번호
설명
가격

4-Methyl-2-pentanone

analytical standard

Hexanal

analytical standard

Dimethyl disulfide

analytical standard

CONDITIONS

sample/matrix

3 g of 2% milk + 10 μL internal standard solution (20 μg/mL 4-methyl-2-pentanone) (9 mL GC vial)

SPME fiber

PDMS/Carboxen, 75 μm (57318)

extraction

headspace, 15 min with constant stirring at 45 °C

desorption process

5 min, 250 °C

column

Supel-Q PLOT, 30 m × 0.32 mm I.D. (24242)

oven

70 °C (2 min) to 140 °C at 6 °C/min (2 min hold) then to 220 °C at 6 °C/min (5 min hold)

inj. temp.

250 °C

detector

GC-MS ion trap, m/z = 33-300

carrier gas

helium, 35 cm/sec

injection

splitless, closed 2 min

liner

0.75 mm liner

sample

Milk

설명

분석 메모

In this study, SPME was used for the analysis of volatiles in milk using a headspace method. Light can catalyze the formation of peroxides that can cleave double bonds in fatty acids. The cleaved by-products form aldehydes that can give the food a pungent off flavor. Dr. Ray Marsili investigated the feasibility of using SPME as a substitute sample preparation device prior to GC/MS analysis. The Carboxen/PDMS fiber was evaluated for the extraction of low ppb levels of pentanal, hexanal, and dimethyldisulfide. The results of the comparison of SPME to dynamic headspace analysis indicate that SPME-GC/MS is more accurate and precise while providing sensitivities equivalent to DH-GC/MS. The GC column used, Supel-Q PLOT, operates via gas-solid chromatography with a DVB adsorbent and permits the separation of a wide range of analytes.