GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Salmon on SPB®-608 (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel QuE Z-Sep, Fast GC Analysis

GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Salmon on SPB®-608 (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel™ QuE Z-Sep, Fast GC Analysis suitable for GC, application for SPE

Materials

SPE 튜브 또는 플레이트

제품 번호
설명
가격

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

Supel QuE

Non-Buffered Tube 1, pk of 50

Supel QuE

Z-Sep+ Bulk, pkg of 20 g

CONDITIONS

sample preparation

Dispersive (QuEChERS)

sample/matrix

raw salmon spiked with PAHs at 100 ng/g

extraction process

add 25 mL acetonitrile and the contents of the Supel QuE Non-buffered tube (55294-U); shake for 5 minutes; centrifuge for 5 minutes at 3200 rpm

extraction tube

Supel QuE Non-buffered tube, magnesium sulfate/sodium chloride (55294-U)

clean-up tube

Supel QuE Z-Sep (500 mg, part no. 55299-U), magnesium sulfate (450 mg, part no. 63135) (55299-U, 63135)

clean-up process

transfer 3 mL of acetonitrile layer, Supel QuE Z-Sep (500 mg) and MgSO4 (450 mg) into an extraction tube; (shake 1 minute; centrifuge for 3 minutes at 3400 rpm; remove 1 mL supernatant for GC-MS analysis)

column

SPB-608, 20 m x 0.18 mm I.D. x 0.18 μm

oven

60 °C (1 min), 25 °C/min to 275 °C, 10 °C/min to 300 °C (13 min)

inj. temp.

265 °C

detector

MS, SIM

MSD interface

300 °C

carrier gas

helium, 1.5 mL/min, constant

injection

1 μL, splitless (1 min)

liner

4 mm I.D. FocusLiner with taper

설명

일반 설명

Stenerson, K.; Ye, M.; Halpenny, M., New Analytical Tools for the Determination of Persistent Organic Pollutants (POPs) in Fatty Food and Beverage Matrices Using QuEChERS Extraction/Cleanup and Gas Chromatography (GC) Analysis. Pittcon 2013. Sigma-Aldrich/Supelco publication T413034

애플리케이션

Extraction solvent: acetonitrile, 25 mL
Clean-up volume: 3 mL
Clearn-up tube: Supel QuE Z-Sep (500 mg, part no. 55299-U, magnesium sulfate (450 mg, part no. 63135)

분석 메모

Sample pretreatment: Place 5 g ground sample into 50 mL centrifuge tube. Spike with 100 ng/g target PAHs and internal standards. Let sit 30 min. Add 5 mL water and homogenize.
This application demonstrates the use of dispersive SPE or QuEChERS for the extraction and cleanup of raw salmon for GC-MS analysis of polynuclear aromatic hydrocarbons (PAHs). Supel QuE Z-Sep sorbent was used in the cleanup step and found to reduce background and yield recoveries >80% for a majority of the PAHs tested.