GC Analysis of PCBs in Whole Cow Milk (Spiked) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel QuE Z-Sep, Fast GC Analysis

GC Analysis of PCBs in Whole Cow Milk (Spiked) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel™ QuE Z-Sep, Fast GC Analysis suitable for GC, application for SPE

Materials

SPE 튜브 또는 플레이트

제품 번호
설명
가격

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

Supel QuE

Non-Buffered Tube 1, pk of 50

Supel QuE

Z-Sep+ Bulk, pkg of 20 g

분석 컬럼

제품 번호
설명
가격

SLB®-5ms Capillary GC Column

L × I.D. 20 m × 0.18 mm, df 0.18 μm

표준

제품 번호
설명
가격

PCB standard solution 7

analytical standard, 10 μg/mL in isooctane

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

10 g whole cow′s milk

extraction tube

Supel QuE Non-buffered tube (4 g magnesium sulfate, 1 g sodium chloride) (55294-U)

clean-up tube

Supel QuE Z-Sep 150 mg, magnesium sulfate (150 mg, part no. 63135) (55299-U)

column

SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)

oven

75 °C (1 min), 12 °C/min to 340 °C (10 min)

inj. temp.

250 °C

detector

μ-ECD, 340 °C

carrier gas

hydrogen, 1.2 mL/min, constant

injection

1 μL splitless (0.75 min)

liner

4 mm I.D. FocusLiner with taper

설명

일반 설명

Reference: Stenerson, K.; Ye, M.; Halpenny, M., New Analytical Tools for the Determination of Persistent Organic Pollutants (POPs) in Fatty Food and Beverage Matrices Using QuEChERS Extraction/Cleanup and Gas Chromatography (GC) Analysis. Pittcon 2013. Sigma-Aldrich/Supelco publication T413034

분석 메모

This application demonstrates the use of dispersive SPE or "QuEChERS" for the extraction and cleanup of whole cow′s milk for GC-ECD analysis of polychlorinated biphenyl (PCBs). Z-Sep sorbent was used in the cleanup step and found to reduce background and yield recoveries >70% for all PCB congeners evaluated except no. 209, which was recovered at 68%.
Extraction solvent: acetone:hexane (1:1), 20 mL
Clean-up volume: 1 mL