콘텐츠로 건너뛰기
Merck

GC Analysis of Amphetamines in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber

GC Analysis of Amphetamines in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materials

SPME 파이버

제품 번호
설명
가격

SPME fiber assembly Polydimethylsiloxane (PDMS)

df 100 μm(PDMS, needle size 24 ga, for use with manual holder

CONDITIONS

sample/matrix

1 mL urine + 0.7g potassium carbonate in 20 mL headspace vial, equilibrated at 80°C, 30 min

SPME fiber

polydimethylsiloxane, 100 μm (57300-U)

derivatization

methyl bis-trifluoroacetamide (headspace, 0.5 min, ambient)

extraction

headspace, 3-5 min, 80 °C

desorption process

1 min, 270 °C

column

100% methyl polysiloxane, 12.5 m × 0.2 mm I.D., 0.33 μm (Supelco equivalent, Equity-1, 28041-U)

oven

60 °C (1 min) to 140 °C (4 min) at 30 °C/min, then to 276 °C at 20 °C/min, 4 min

inj. temp.

splitless (closed 1 min), 270 °C

detector

MS, full scan