콘텐츠로 건너뛰기
Merck
모든 사진(1)

문서

G005902

HPLC Analysis of Sugars by HILIC Chromatography using Ascentis® Express OH-5 Column with ELSD Detection

application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable

검사 파라미터

column: Ascentis Express OH-5, 15 cm x 4.6 mm, I.D., 2.7 μm particles (53778-U)
mobile phase: [A] 0.005 M ammonium formate, pH 3.15 with concentrated TFA [B] acetonitrile (30:70, A:B)
flow rate: 0.5 mL/min
pressure: 841 psi (58 bar)
column temp.: 35 °C
detector: ELSD (detector other: Nebulizer 60º C, Evaporator 80º C, Gas flow 1.60 SLM (Standard Liters/ Minute), LED 100%, Smoothing 5.0 S, PMT ( Photomultiplier) Gain 1.0)
injection: 10 μL
sample: 0.5 mg/mL each prepared in 50:50 Water:Acetonitrile

적합성

application for HPLC

응용 분야

food and beverages

분석 메모

This application shows the analysis of sugars by HILIC Chromatography with ELSD detection. The purpose of the work shown here was to develop a HILIC method that could quickly separate mono- and di-saccharides with good sensitivity and reproducibility utilizing the Ascentis Express OH-5 column and an Evaporative Light Scattering Detecctor (ELSD) detector.

법적 정보

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Dextrose, Pharmaceutical Secondary Standard; Certified Reference Material
    Sucrose, Pharmaceutical Secondary Standard; Certified Reference Material
    Maltitol, Pharmaceutical Secondary Standard; Certified Reference Material

가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

죄송합니다. 지금은 이 제품에 대한 COA이(가) 온라인에서 제공되지 않습니다.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..

고객지원팀으로 연락바랍니다.