콘텐츠로 건너뛰기
Merck
모든 사진(1)

문서

G005894

HPLC Analysis of Pain Relievers on Ascentis® Express C18

application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable

검사 파라미터

column: Ascentis Express C18, 10 cm x 4.6 mm I.D., 5 μm particles (50536-U)
mobile phase: [A] water with 3% acetic acid; [B] methanol with 3% acetic acid; (79:21, A:B)
flow rate: 2.0 mL/min
pressure: 3290 psi (227 bar)
column temp.: 45 °C
detector: UV,275 nm
injection: 10 μL
sample: 100 - 360 μg/mLin mobile phase

적합성

application for HPLC

응용 분야

clinical
pharmaceutical (small molecule)

분석 메모

This application shows the improvements in efficiency that may be obtained for USP methods using modern Fused-Core technology.

법적 정보

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Acetaminophen, meets USP testing specifications, 98.0-102.0%, powder
    Caffeine, meets USP testing specifications, anhydrous
    Benzoic acid, meets USP testing specifications
    Aspirin, meets USP testing specifications

Choose from one of the most recent versions:

시험 성적서(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..

고객지원팀으로 연락바랍니다.