콘텐츠로 건너뛰기
Merck
모든 사진(1)

문서

G005723

LC/MS Analysis of Polypeptide Drugs on Ascentis® Express RP-Amide

application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable

검사 파라미터

column: Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2.7 μm particles (53913-U)
mobile phase: [A] 10 mM ammonium acetate in water, pH 4.2; [B] 10 mM ammonium acetate in 90:10 (v/v) acetonitrile:water, pH 4.7
gradient: held at 20% B for 0.1 min; 20 to 80% B in 2.9 min; 80 to 20% B in 0.01 min; held at 20% B for 0.99 min
flow rate: 0.3 mL/min
column temp.: 35 °C
detector: TOF, ESI(+), m/z 100-2500
injection: 3 μL
sample: 1 μL/mL in 10mM ammonium formate, 80:20, acetonitrile:water

적합성

application for HPLC

응용 분야

life science and biopharma

분석 메모

This application demonstrates the suitability of the Ascentis Express RP Amide for the analysis of polypeptide drugs.

법적 정보

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Oxytocin
    [deamino-Cys1, D-Arg8]-Vasopressin acetate salt hydrate

Choose from one of the most recent versions:

시험 성적서(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..