GC Analysis of Aromatics in Reformulated Gasoline on SLB®-IL111

suitable for GC

로그인조직 및 계약 가격 보기


About This Item

기술

gas chromatography (GC): suitable

검사 파라미터

column: SLB-IL111, 60 m x 0.25 mm I.D., 0.20 μm (28928-U)
oven: 50 °C (3 min), 15 °C/min to 265 °C (5 min)
inj. temp.: 250 °C
detector: FID, 265 °C
carrier gas: helium, 30 cm/sec
injection: 0.5 μL, 200:1 split
liner: 4 mm I.D., splitless type, single taper wool packed FocusLiner design
sample: reformulated unleaded gasoline

응용 분야

petroleum

분석 메모

This application illustrates the analysis of a sample of reformulated gasoline on a 60 m SLB-IL111 ionic liquid column. Due to the extremely polar nature of the phase, the aliphatics elute prior to the aromatics. The high level of ethanol in reformulated gasoline can pose a problem in the detection and quantification of benzene. The SLB-IL111 provided separation between these two compounds, with a starting temperature of 50 °C providing the best resolution. The SLB-IL111 has a maximum operating temperature of 270 °C, which is extremely high for a phase of this polarity. The high maximum temperature was utilized in this analysis to elute the napthalenes in < 15 minutes.

법적 정보

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Ethanol
    Dodecane
    Benzene
    Tridecane
    Toluene
    Tetradecane
    Ethylbenzene
    p-Xylene
    m-Xylene
    o-Xylene
    1,2,4-Trimethylbenzene
    1,2,3-Trimethylbenzene
    Durene
    Naphthalene
    2-Methylnaphthalene
    1-Methylnaphthalene
    모두 보기 (16)

분석 컬럼


가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 문서 section.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..

고객지원팀으로 연락바랍니다.