콘텐츠로 건너뛰기
Merck
모든 사진(1)

주요 문서

G003777

GC Analysis of Triazine Herbicides on SLB®-5ms after SPME using 60 μm Carbowax® Fiber

application for SPME, suitable for GC

로그인조직 및 계약 가격 보기


About This Item

기술

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

검사 파라미터

sample/matrix: Triazine herbicides at 2 ppb in 0.1M phosphate buffer, pH 7, with 25% NaCl, 9 mL in 10 mL screw cap vial
SPME fiber: 60 μM Carbowax (PEG) (57354-U)
extraction: 30 min direct immersion with agitation using CombiPal program
desorption process: 2 min at 250°C, splitless for 0.75 min then opened
column: SLB-5ms, 30 m x 0.25 mm I.D., 0.5μm film
oven: 60°C(1 min) to 150°C @ 25°C/min to 230°C @ 5°C/min (0.5 min)
inj. temp.: Splitless/split with 0.75mm ID liner and Merlin Microseal
detector: ITMS m/z=150-250 @ 0.5 μsec per scan, transfer line 310°C
carrier gas: Helium, 1mL/min constant flow

적합성

application for SPME

응용 분야

agriculture
chemicals and industrial polymers
environmental
food and beverages

법적 정보

Carbowax is a registered trademark of Union Carbide
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Prometon
    Simazine
    Atrazine
    Propazine
    Simetryn
    Ametryn
    Prometryn
    Terbutryn
    Cyanazine
    모두 보기 (9)