콘텐츠로 건너뛰기
Merck
모든 사진(1)

문서

G003764

HPLC Analysis of Furosemide (Lasix) in Horse Serum on Discovery® C18 after SPE using Discovery® DSC-18

application for SPE, application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable
solid phase extraction (SPE): suitable

검사 파라미터

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: acidified horse serum spiked with indapamide at 10 mg/mL
SPE tube/cartridge: Discovery DSC-18, 50 mg/1 mL (52601-U)
condition: 1 mL methanol; 1 mL 10 mM potassium phosphate, pH 3
sample addition: 1 mL
washing: 10 mM potassium phosphate, pH 3
elution: 1 mL 60% methanol in DI water.
column: Discovery C18, 15 cm x 4.6 mm I.D., 5 μm particles (504955)
mobile phase: [A] 10 mM KH2PO4, pH 3 (adjusted with H3PO4): [B] acetonitrile; (60:40, A:B)
flow rate: 1 mL/min
column temp.: 35 °C
detector: UV, 234 nm
injection: 10 μL

적합성

application for HPLC
application for SPE

응용 분야

forensics and toxicology

애플리케이션

Sample Pre-treatment:
Neat horse serum was acidified with 10 mL 6 M HCl/mL serum. Internal standard (I.S.) blank was prepared by spiking acidified horse serum with indapamide at 10 mg/mL. Sample A = 10 mg/mL furosemide in I.S. blank. Sample B = 5 mg/mL furosemide in I.S. blank. Sample C = 0.5 mg/mL furosemide in I.S. blank. Sample D = 0.1 mg/mL furosemide in I.S. blank. Sample E = 0.05 mg/mL furosemide in I.S. blank.

법적 정보

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany

SPE 튜브 또는 플레이트

분석 컬럼


가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

죄송합니다. 지금은 이 제품에 대한 COA이(가) 온라인에서 제공되지 않습니다.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..