- Instrumental planar chromatographic method for determination of carbamazepine in human serum.
Instrumental planar chromatographic method for determination of carbamazepine in human serum.
An instrumental planar chromatographic (HPTLC) method for quantification of carbamazepine in human serum was developed using liquid-liquid extraction with dichloromethane, fluorescence activation with perchloric acid 60%/ethanol/water (1:1:1, v/v) and fluorescence detection. Planar chromatographic separation was performed on precoated silica gel F254 HPTLC plates using a mixture of ethyl acetate/toluene/methanol/acetic acid glacial (5:4:0.5:0.5, v/v) as mobile phase. Densitometric detection was done at 366 nm. The method was validated for linearity, precision and accuracy. Linear calibration curves in the range of 3 and 20 ng/microL showed correlation coefficient of 0.998. The intra-assay and inter-assay precision, expressed as the RSD, were in the range of 0.41-1.24% (n = 3) and 2.17-3.17% (n = 9), respectively. The LOD was 0.19 ng, and the LOQ was 0.57 ng. Accuracy, calculated as percentage recovery, was between 98.98 and 101.96%, with a RSD not higher than 1.52%. The method was selective for the active principle tested. In conclusion, the method is useful for quantitative determination of carbamazepine in human serum.