796-0376
OSHA Method 42/47 and ASTM® D5836: HPLC Analysis of Isocyanates on SUPELCOSIL™ LC-8 after Collection/Desorption using ORBO™-80
application for air monitoring, application for HPLC
About This Item
Recommended Products
technique(s)
HPLC: suitable
test parameters
sample/matrix: calibration standard: isocyanate derivatives, 5 μg/mL in acetonitrile blank: 1-(2-pyridyl)piperazine (1-2PP) coated glass fiber filter desorbed in acetonitrile/DMSO, 9:1
adsorbent tube: ORBO-80 (20811)
column: SUPELCOSIL LC-8, 25 cm × 4.6 mm I.D., 5 μm particles (58297)
mobile phase: 0.05 M ammonium acetate in water:acetonitrile, 70:30 (pH 6-6.2 with acetic acid)
flow rate: linear gradient, 0.7 mL/min to 2 mL/min in 15 min
detector: UV, 254 nm
injection: 10 μL
suitability
application for HPLC
application for air monitoring
application(s)
air monitoring
industrial hygiene
Legal Information
Analyte
analytical column
related product
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service