Gas chromatography-optical fiber detector for the speciation of aromatic hydrocarbons in confined areas.

Analytical sciences : the international journal of the Japan Society for Analytical Chemistry (2008-08-12)
Lurdes I B Silva, Teresa A P Rocha-Santos, A C Duarte
ABSTRACT

An analytical method, based on separation with gas chromatography (GC) and detection with optical fiber (OF), was used for the separation, detection and quantification of benzene, toluene, ethylbenzene, p-xylene, m-xylene and o-xylene. The use of OF as a detector is based on the variations of the reflected optical power detected when the aromatic compounds eluted from the GC column are sorbed in a thin polymeric film on a single-mode OF. General figures of merit, such as the analytical time, analytical error and analytical performance of GC-OF were similar to those of the classical analytical methods, such as a gas chromatography-flame ionization detector (GC-FID). However, the developed GC-OF method constitutes a much less expensive alternative for the speciation of aromatic hydrocarbons compounds, with high accuracy, and being most suitable for actual monitoring work on confined environments.

MATERIALI
N° Catalogo
Marchio
Descrizione del prodotto

Supelco
cartuccia ORBO 32 piccola, con carbone attivo da cocco, (20/40), 100/50 mg, W,F,F separators, O.D. × L 6 mm × 75 mm, pkg of 50 ea
Supelco
m-Xylene, analytical standard
Supelco
m-Xylene, Pharmaceutical Secondary Standard; Certified Reference Material