The methylene blue dye on the strip undergoes a chemical reduction reaction in the absence of oxygen, turning colorless to indicate a total anaerobic atmosphere. The reaction zone on the test strip returns to normal (blue color) upon re-introduction of oxygen in the environment.
1.32371
Anaerotest™ Strips
Test strips to indicate an anaerobic atmosphere., pkg of 50 strips per pack
Synonym(s):
Anaerobic indicator strip
About This Item
Recommended Products
Quality Level
form
solid
packaging
pkg of 50 strips per pack
application(s)
pharmaceutical
compatibility
Anaerocult™ system
storage temp.
15-25°C
Related Categories
General description
Application
Legal Information
related product
Storage Class Code
11 - Combustible Solids
WGK
WGK 3
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Articles
Anaerocult® system supports reliable anaerobe cultivation for microbiological research needs.
Anaerocult® system supports reliable anaerobe cultivation for microbiological research needs.
Anaerocult® system supports reliable anaerobe cultivation for microbiological research needs.
Anaerocult® system supports reliable anaerobe cultivation for microbiological research needs.
-
Once the strip turns white in anaerobic conditions, if you place it in aerobic conditions, how long does it take for the strip to turn back to blue? In other words, is there a time delay between O2 exposure and the time the strip turns blue?
1 answer-
Helpful?
-
-
What is the limit of resudial Oxygen concetration in the Antmosphere to indicate the white colour. And what oxygen concetration makes it back to the blue colour. What are storage conditions of and expiry time of an opned bag of test strips.
1 answer-
The anaerotest strips 1.32371.0001 have a detection limit close to 0.5% O2 in the atmosphere. It will change color when crossing that threshold.
Helpful?
-
-
Wie lange kann ich mit einem Streifen eine Umgebung monitoren? Wann muss der Streifen ausgetauscht werden?
1 answer-
Reference 1.32371.0001 Anaerotest strip indicator has no validated maximum duration after activation. We know that the product works for the amount of time needed for its specific application: it will be able to revert color as long as day 5 after activation.
Helpful?
-
-
I placed a strip inside an anaerobic chamber to check O2 presence. It turned white in about 13 minutes, but after 16 hrs approximately, it turned back blue. I kept it there for another 24 hrs, and it turned white again. Is this normal?
1 answer-
Anaerotest™ strips are designed to indicate an anaerobic atmosphere via a color change on the test strip reaction zone. The methylene blue dye in the absence of oxygen undergoes a chemical reduction reaction and turns colorless thereby, indicating a total anaerobic atmosphere. The reaction zone on the test strip returns to normal (blue color) upon re-introduction of oxygen in the environment. During the incubation, the aspect of the test strips could be regularly inspected to check if the anaerobic conditions are still met. If the test strips become blue again during incubation it could indicate a leak in the incubation jar.