794-0399
GC Analysis of a 20-Component Organochlorine Pesticide Mix in Water on a 5% Phenyl Column after SPE using Supelclean™ ENVI™-8
suitable for GC, application for SPE
Synonym(s):
Chlorinated pesticides in hazardous waste
About This Item
technique(s)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: waste water spiked with pesticides
SPE tube/cartridge: Supelclean ENVI-8, glass hardware, 500 mg/6 mL (57107 (500 mg/6 mL); 57108 (1 g/6 mL))
condition: 3 mL methanol; h 2 mL 5% methanol in water
sample addition: load sample at 5 mL/min
drying: 2-3 min with nitrogen stream
elution: 2 x 4 mL hexane:acetone (90:10), drop wise. allow to soak into bed
column: 5% Phenyl Polysiloxane, 15 m × 0.53 mm I.D., 0.50 μm (Supelco equivalent, Equity-5, 28252-U)
oven: 150 °C (0.5 min) to 275 °C (5 min)
inj. temp.: 200 °C
detector: ECD, 300 °C
carrier gas: helium
injection: 1 μL split/splitless (45 sec delay)
suitability
application for SPE
application(s)
environmental
food and beverages
Application
Add 5 mL methanol to 100 mL aqueous hazardous waste. Adjust to pH 5-7 if necessary.
Legal Information
Analyte
SPE tube or plate
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service