Saltar al contenido
Merck

LC/MS Analysis of Polypeptide Drugs on Ascentis® Express RP-Amide

LC/MS Analysis of Polypeptide Drugs on Ascentis® Express RP-Amide application for HPLC

Materiales

analytical column

Referencia del producto
Descripción
Precios

Ascentis® Express RP-Amide, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

mobile phase component

Referencia del producto
Descripción
Precios

Ammonium acetate

LiChropur, eluent additive for LC-MS

standard

Referencia del producto
Descripción
Precios

Oxytocin

United States Pharmacopeia (USP) Reference Standard

Desmopressin

European Pharmacopoeia (EP) Reference Standard

Eptifibatide acetate

≥98% (HPLC)

Bivalirudin trifluoroacetate salt

≥97% (HPLC)

CONDITIONS

column

Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2.7 μm particles (53913-U)

mobile phase

[A] 10 mM ammonium acetate in water, pH 4.2; [B] 10 mM ammonium acetate in 90:10 (v/v) acetonitrile:water, pH 4.7

gradient

held at 20% B for 0.1 min; 20 to 80% B in 2.9 min; 80 to 20% B in 0.01 min; held at 20% B for 0.99 min

flow rate

0.3 mL/min

column temp.

35 °C

detector

TOF, ESI(+), m/z 100-2500

injection

3 μL

sample

1 μL/mL in 10mM ammonium formate, 80:20, acetonitrile:water

Descripción

Analysis Note

This application demonstrates the suitability of the Ascentis Express RP Amide for the analysis of polypeptide drugs.

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany