Saltar al contenido
Merck

US EPA Method 608.1/608.2: GC Analysis of Organochlorine Pesticides on Equity-1701

US EPA Method 608.1/608.2: GC Analysis of Organochlorine Pesticides on Equity-1701 suitable for GC

Materiales

Analytical column

Referencia del producto
Descripción
Precios

SPB®-608

L × I.D. 30 m × 0.25 mm, df 0.25 μm

SPB®-608

L × I.D. 30 m × 0.53 mm, df 0.50

Columna para GC capilar Equity -1701

L × I.D. 30 m × 0.25 mm, df 0.25 μm

Columna para GC capilar Equity -1701

L × I.D. 30 m × 0.25 mm, df 1.00 μm

Columna para GC capilar Equity -1701

L × I.D. 30 m × 0.53 mm, df 0.50 μm

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.00 μm

SLB®-5ms Capillary GC Column

L × I.D. 15 m × 0.10 mm, df 0.10 μm

For use with

Referencia del producto
Descripción
Precios

Wheaton wide-mouth bottle

amber soda-lime glass bottle, capacity 1000 mL, white polypropylene cap, (PTFE faced LDPE liner), case of 12 ea

Optional

Referencia del producto
Descripción
Precios

Polyimide Sealing Resin

pkg of 5 g

guard column

Referencia del producto
Descripción
Precios

Nonpolar Fused Silica Tubing

L × I.D. 5.0 m × 0.10 mm

Nonpolar Fused Silica Tubing

L × I.D. 5.0 m × 0.25 mm

Nonpolar Fused Silica Tubing

L × I.D. 5.0 m × 0.53 mm

required but not provided

Referencia del producto
Descripción
Precios

"Y" GlasSeal Connector, Borosilicate Glass

pkg of 1 ea

"Y" Glassseal Connectors

pkg of 1 ea

"Y" Glassseal Connectors

pkg of 3 ea

CONDITIONS

column

Equity-1701, 30 m x 0.25 mm I.D., 0.25 μm (28372-U)

oven

75 °C (2 min), 15 °C/min to 280 °C (5 min)

inj. temp.

250 °C

detector

ECD, 300 °C

carrier gas

helium, 1.2 mL/min. constant

injection

1 μL, splitless (splitter open after 0.75 min)

liner

4 mm I.D., single taper

sample

mixture of 608.1 and 608.2 calibration solutions (40351-U and 40352-U) diluted to 100 ppb in n-hexane

Descripción

Legal Information

null