796-0376
OSHA Method 42/47 and ASTM® D5836: HPLC Analysis of Isocyanates on SUPELCOSIL™ LC-8 after Collection/Desorption using ORBO™-80
application for air monitoring, application for HPLC
About This Item
Productos recomendados
technique(s)
HPLC: suitable
test parameters
sample/matrix: calibration standard: isocyanate derivatives, 5 μg/mL in acetonitrile blank: 1-(2-pyridyl)piperazine (1-2PP) coated glass fiber filter desorbed in acetonitrile/DMSO, 9:1
adsorbent tube: ORBO-80 (20811)
column: SUPELCOSIL LC-8, 25 cm × 4.6 mm I.D., 5 μm particles (58297)
mobile phase: 0.05 M ammonium acetate in water:acetonitrile, 70:30 (pH 6-6.2 with acetic acid)
flow rate: linear gradient, 0.7 mL/min to 2 mL/min in 15 min
detector: UV, 254 nm
injection: 10 μL
suitability
application for HPLC
application for air monitoring
application(s)
air monitoring
industrial hygiene
Legal Information
Analito
analytical column
related product
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico