TSKgel® IC-Cation SW HPLC Column, phase sulfonic acid, L × I.D. 5 cm × 4.6 mm, 5 μm particle size
material
stainless steel
Agency
suitable for USP L52
product line
TSKgel®
manufacturer/tradename
Tosoh 08055
parameter
1.2 mL/min flow rate 10-45 °C temperature 70 bar max. pressure
technique(s)
HPLC: suitable
L × I.D.
5 cm × 4.6 mm
matrix
spherical silica
matrix active group
sulfonic acid phase
particle size
5 μm
operating pH
2.0-7.5
separation technique
cation exchange
General description
Ion chromatography (IC) is a method of analyzing inorganic and organic ions using an HPLC system in which charged species are separated on a anion or cation exchange column and detected by a conductivity detector. In many cases IC has superseded conventional wet chemistry methods such as titration, photometry, gravimetry, and colorimetry.
TSKgel® IC-Cation SW is a high performance ion chromatography column used without a suppression system. Typical applications include the analysis of transition metal ions, such as Fe, Cu, Ni, Zn, Mn, etc. using ethylene-diamine/citric acid or ethylene-diamine/tartaric acid as eluent, and the analysis of amine compounds which have a strong hydrophobic group such as R4N+or R3N, in which R is an alkyl group such as butyl, octyl, etc., using aq. nitric acid/acetonitrile as eluent.
Recommended products
Discover LiChropur reagents ideal for HPLC or LC-MS analysis
Legal Information
TSKgel is a registered trademark of Tosoh Corporation
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Lot/Batch Number
Sorry, we don't have COAs for this product available online at this time.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.