Typically 3 years from sterilization.
Seleccione un Tamaño
| Talla/SKU | Disponibilidad | Precio |
|---|
Acerca de este artículo
Quality Segment
material
polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing
product line
Stericup®
feature
hydrophilic
packaging
pack of 12 ea
greener alternative product characteristics
Waste Prevention
Learn more about the Principles of Green Chemistry.
sustainability
Greener Alternative Product
parameter
max. temp. at 45 °C
description
2.4 mL (hold-up volume)
technique(s)
sterile filtration: suitable
H
193 mm
diam.
9.3 cm
filter capacity
500 mL
filter diam.
73 mm
filtration area
40 cm2
matrix
Express PLUS™
pore size
0.22 μm pore size
fitting
inlet thread(s), 38 mm (1 1/2 in.) thread
greener alternative category
shipped in
ambient
storage temp.
no temp limit
General description
Para obtener más información, visite la BASE DE DATOS ACT
Los filtros esterilizantes Stericup® E y Steritop® E están diseñados como una Alternativa más Ecológica para la esterilización mediante filtración al vacío que reduce de manera significativa el plástico y los envases.
Este Stericup® E es un collarín de rosca de 38 mm que tiene una membrana de filtración de polietersulfona (PES) con un tamaño de poro de 0,22 μm combinada con un matraz receptor de 500 ml y una tapa para el procesamiento y almacenamiento. La rosca de 38 mm se conecta directamente a nuestras botellas de medios y a la mayoría de otras botellas de medios comerciales estándar.
Use solo botellas de vidrio o plástico diseñadas para aplicaciones al vacío.
Application
- Medios de cultivo de tejidos +/- aditivos
- Amortiguadores
- Disoluciones biológicas
Features and Benefits
- El collar del filtro se desconecta de la botella receptora con solo un cuarto de vuelta—reduciendo así la probabilidad de derrame y haciendo que sea más fácil su manipulación sobre la mesa de trabajo o debajo de la campana.
- La identificación del producto en negrita en el collarín del filtro muestra claramente las características de la membrana, como su composición y el tamaño de los poros, para una selección de productos rápida y apropiada para la aplicación.
- El tapón de cierre mediante un clic está bridado y modificado para permitir un fuerte agarre en condiciones húmedas o secas—y un tope táctil confirma un cierre seguro. Preocúpese menos de la posibilidad de derramamiento o contaminación de los contenidos esterilizados.
- Una amplia superficie esmerilada de escritura en la botella receptora y el color más claro de la tapa facilita el etiquetado claro para mejorar la legibilidad y reducir los frotis ilegibles.
- La tapa descansa sobre el lateral para evitar el riesgo de contaminación.
- Las botellas receptoras son apilables para ahorrar espacio de almacenamiento.
Legal Information
1 of 1
Este artículo | |||
|---|---|---|---|
| description pore size 0.22 μm, polyethersulfone membrane, filter capacity 500 mL, 38 mm (1 1/2 in.) thread, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, capacity 500 mL, 45 mm thread, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 1000 mL, 38 mm (1 1/2 in.) thread, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 1000 mL, 45 mm thread, pack of 12 ea |
| pore size 0.22 μm pore size | pore size 0.22 μm pore size | pore size 0.22 μm pore size | pore size 0.22 μm pore size |
| filter capacity 500 mL | filter capacity 500 mL | filter capacity 1000 mL | filter capacity 1000 mL |
| material polystyrene funnel(s), polyethersulfone membrane, polystyrene bottle, polystyrene housing | material polystyrene bottle, polystyrene housing, polyethersulfone membrane, polystyrene funnel(s) | material polyethersulfone membrane, polystyrene funnel(s), polystyrene bottle, polystyrene housing | material polystyrene funnel(s), polystyrene housing, polyethersulfone membrane, polystyrene bottle |
| feature hydrophilic | feature hydrophilic | feature hydrophilic | feature hydrophilic |
| packaging pack of 12 ea | packaging pack of 12 ea | packaging pack of 12 ea | packaging pack of 12 ea |
| fitting 38 mm (1 1/2 in.) thread, inlet thread(s) | fitting inlet thread(s), 45 mm thread | fitting inlet thread(s), 38 mm (1 1/2 in.) thread | fitting inlet thread(s), 45 mm thread |
Still not finding the right product?
Explore all of our products under Unidad de filtración Stericup® E-GP ecológica
Elija entre una de las versiones más recientes:
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
-
What is the expiration date of Stericup® and Steritop® filters?
1 answer-
Helpful?
-
-
Are Stericup® and Steritop® filters GMP-certified?
1 answer-
No, Stericup® and Steritop® devices are not GMP-approved.
Helpful?
-
-
Can Stericup® and Steritop® filters be used to prepare solutions that will be administered to patients?
1 answer-
No, all Stericup® and Steritop® devices are Research Use Only.
Helpful?
-
-
What Bacterial Endotoxins Test is used?
1 answer-
The test method used complies with USP 85 and EP 2.6.14 to ensure the absence of endotoxin in our products.
Helpful?
-
-
What tubing can be used with Stericup® and Steritop® vacuum filters?
What tubing can be used with Stericup® and Steritop® vacuum filters?
1 answer-
Stericup® and Steritop® filters are fitted with a 1/4” tube adapter. We recommend 3/16 in. (4.8 mm) ID silicone tubing to fit this tube adapter: https://www.sigmaaldrich.com/product/mm/xx7100004
Helpful?
-
-
What leachables are detected during filtration processes using Stericup® filters?
1 answer-
We have not performed extractables/leachables studies for our Stericup® filters.
Helpful?
-
-
The User Guide says “Use only glass or plastic bottles designed for vacuum applications”. Can you specify which bottles are compatible?
1 answer-
Steritop® Quick Release bottle top units require a vacuum-rated receiver bottle with either a GL33 or GL45 mm closure thread as specified. Steritop® E bottle top units require a vacuum-rated receiver bottle with a GL45 mm closure thread. Bottles no larger than 2 L should be used.
Glass bottles, GL33:
• Wheaton media bottles (glass), 500 mL (Product Z364886)
• Wheaton media bottles (glass), 1 L (Product Z364894)
Glass bottles, GL45
• Corning® glass media storage bottle, 125 mL (Product CLS1399125)
• Corning® glass media storage bottle, 250 mL (Product CLS1399250)
• Corning® glass media storage bottle, 500 mL (Product CLS1399500)
• Corning® glass media storage bottle, 1 L (Product CLS13991L)
• Pyrex® round media storage bottle, 500 mL (Product CLS1395500)
• Pyrex® round media storage bottle, 1 L (Product CLS13951L)
• KIMBLE® KIMAX® Media bottle, 250 mL (Product DWK14395250)
• KIMBLE® KIMAX® Media bottle, 500 mL (Product DWK14395500)
• KIMBLE® KIMAX® Media bottle, 1 L (Product DWK143951000)
Plastic (pre-sterile) bottles, GL45:
• Stericup® Quick Release Receiver flask, 250 mL (Product S200B02RE)
• Stericup® Quick Release Receiver flask, 500 mL (Product S200B05RE)
• Stericup® Quick Release Receiver flask, 1000 mL (Product S200B10RE)Helpful?
-
-
What is the difference between classic Stericup® filters and Stericup® “Quick Release” filters?
1 answer-
Stericup® Quick Release, or “QR”, filters feature a connection between funnel filter and receiver bottle that requires only a quarter-turn release to facilitate funnel removal. These filters come fully engaged – there is no need to tighten before use. The receiver bottle cap has flanges on the top surface to improve grip, as well as a positive lock stop to ensure a leak-proof seal with the receiver bottle. Other improvements include a large, frosted writing surface on the receiver bottle and a lighter cap color to facilitate labelling, more prominent volume markings on the filtration funnel and receiver bottle, and a product label with a peel-off lot record strip on the bag for easy viewing in storage.
Helpful?
-
-
Can we filter oil-type liquids with Stericup® filters?
Can we filter oil-type liquids with Stericup® filters?
1 answer-
Likely not, due to pore size/possible fouling/chemical compatibility. Refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
Helpful?
-
-
What is the maximum vacuum pressure for filtration?
1 answer-
700 mm Hg (27.5” Hg, 0.9 bar) differential vacuum at 25 °C (77 °F)
Helpful?
-