Accéder au contenu
Merck

HPLC Analysis of Alkaloids from Herbal Supplements on Ascentis® Express F5, HILIC Mode

HPLC Analysis of Alkaloids from Herbal Supplements on Ascentis® Express F5, HILIC Mode application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Ascentis® Express F5, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 3 cm × 3 mm

CONDITIONS

column

Ascentis® Express F5, 3 cm x 3.0 mm I.D., 2.7 μm particles (53574-U)

mobile phase

2 mM ammonium acetate in 10:90 water:acetonitrile, pH to 6.0 with acetic acid

flow rate

0.5 mL/min

pressure

1100 psi (76 bar)

column temp.

35 °C

detector

UV, 275 nm

injection

2 μL

sample

methanolic extract of Yohimbe Herbal Supplement