Accéder au contenu
Merck
Toutes les photos(1)

Key Documents

G005740

HPLC Analysis of Spice Cannabinoids on Ascentis® Express C18

application for HPLC

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

HPLC: suitable

Paramètres du test

column: Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)
mobile phase: [A] 5 mM ammonium formate water; [B] 5 mM ammonium formate (95:5, acetonitrile:water)
gradient: 68 to 100% B in 8 min; held at 100% B for 2 min
flow rate: 0.6 mL/min
pressure: 4351 psi (300 bar)
column temp.: 30 °C
detector: UV, 265 nm
injection: 3 μL

Adéquation

application for HPLC

Application(s)

forensics and toxicology

Remarque sur l'analyse

Synthetic cannabinoids (e.g. "Spice") are a type of designer drug used to get a cannabis-type high. New synthetic cannabinoids are continually being introduced as suppliers tweak the molecular structures. The ability to rapidly and reliably identify the continually changing population of these compounds in the blood or urine suspected users is a significant analytical challenge facing forensic chemists. Here is shown the selectivity provided by C18 phase chemistry on an Ascentis Express C18 column with UV detection. Cerilliant CRMs provided reliable identification and quantification.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    JWH-073 N-(3-hydroxybutyl) metabolite solution, 100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    Spice Cannabinoid Mix solution, 100 μg/mL each component in acetonitrile, ampule of 1.0 mL, certified reference material, Cerilliant®
    JWH-073 solution, 100 μg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®
    HU-210 solution, 100 μg/mL in methanol, (Spice Cannabinoid), ampule of 1 mL, certified reference material, Cerilliant®
    JWH-018 solution, 100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Faites votre choix parmi les versions les plus récentes :

Certificats d'analyse (COA)

Lot/Batch Number

Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.

Si vous avez besoin d'assistance, veuillez contacter Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique