G005652
HPLC Analysis of 16 Triazine and Phenyl Urea Compounds on Ascentis® Express RP-Amide, Modified Gradient
application for HPLC
About This Item
Technique(s)
HPLC: suitable
Paramètres du test
column: Ascentis Express RP-Amide, 15 cm x 3 mm I.D., 2.7 μm particles (53919-U)
mobile phase: [A] water; [B] acetonitrile
gradient: 10 to 25% B in 30 min; 25 to 60% B in 15 min; held at 60% B for 5 min
flow rate: 0.4 mL/min
column temp.: 35 °C
detector: UV, 240 nm
injection: 5 μL
sample: 20 mg/L each (except terbuthylazin desethyl and terbuthylazine 60 mg/L) in 60:40, water: acetonitrile
Adéquation
application for HPLC
Application(s)
agriculture
chemicals and industrial polymers
Remarque sur l'analyse
temperature can improve resolution of the critical pair.
Informations légales
Analyte
Colonne analytique
Choose from one of the most recent versions:
Certificats d'analyse (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique