G005293
GC Analysis of Trichloroanisole (TCA) and Precursors on SLB®-5ms after SPME using 100 μm a PDMS Fiber
suitable for GC, application for SPME
About This Item
Technique(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
Paramètres du test
sample/matrix: each analyte at 100 ng/L in a 12% ethanol solution
SPME fiber: metal fiber coated with 100 μm PDMS (57928-U)
extraction: headspace, 50 °C for 30 min, with stirring
desorption process: 250 °C for 3 min
column: SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven: 50 °C (1 min), 25 °C/min to 280 °C
inj. temp.: 250 °C
detector: GC/MS, ECD, 290 °C
carrier gas: helium, 1.5 mL/min, constant
injection: splitless/split, closed 2 min
liner: 0.75 mm I.D. SPME
sample: Wine cork
Adéquation
application for SPME
Application(s)
food and beverages
Remarque sur l'analyse
Informations légales
Analyte
Colonne analytique
Fibre SPME
Faites votre choix parmi les versions les plus récentes :
Certificats d'analyse (COA)
Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.
Si vous avez besoin d'assistance, veuillez contacter Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique