CHVE72HE3
Millipore Express® SHRP High Area, Cartridge
with prefilter, cartridge nominal length 20 in. (50 cm), filter diam. 2.7 in. (6.9 cm), Code 7
About This Item
Matériaux
EPDM rubber seal
polypropylene support
Niveau de qualité
Conformité réglementaire
meets FDA 21 CFR 177-182 (all component materials)
Stérilité
non-sterile
Mode de stérilisation compatible
autoclavable compatible
steam-in-place compatible
Gamme de produits
Millipore Express®
Caractéristiques
hydrophilic
with prefilter
Fabricant/nom de marque
Millipore Express®
Paramètres
≤108.4 mL/min air diffusion at 3.45 bar (50 psig) (in water)
1700 mbar max. differential pressure (25 psi) at 80 °C (Forward)
1700 mbar max. inlet pressure (25 psi) at 80 °C (Forward)
2100 mbar max. inlet pressure (30 psi) at 25 °C (Reverse)
340 mbar max. differential pressure (5 psi) at 135 °C (Forward)
340 mbar max. inlet pressure (5 psi) at 135 °C (Forward)
69 mbar max. inlet pressure (1 psi) at 135 °C (Reverse)
6900 mbar max. differential pressure (100 psi) at 25 °C (Forward)
6900 mbar max. inlet pressure (100 psi) at 25 °C (Forward)
Technique(s)
sterile filtration: suitable
Longueur
20 in.
Largeur
2.8 in.
Longueur nominale des cartouches
20 in. (50 cm)
Diam. du filtre
2.7 in. (6.9 cm)
filter filtration area
2.0 m2
Taille du filtre
20 in.
Impuretés
≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
Matrice
Millipore Express® SHR (w/Prefilter High Area)
pore size
0.5/0.1 μm pore size
Point de bulle
≥2586 mbar (37.5 psig), nitrogen with 70/30% IPA/water
Code de la cartouche
Code 7
Vous recherchez des produits similaires ? Visite Guide de comparaison des produits
Description générale
Conditionnement
Notes préparatoires
This product was manufactured with a MILLIPORE EXPRESS® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
Remarque sur l'analyse
Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2 using ASTM® F838 methodology.
Autres remarques
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer wetting instructions of Filters with MILLIPORE EXPRESS® Membrane user guide
- Storage Statement: Please refer Filters with MILLIPORE EXPRESS® Membrane user guide
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations
Informations légales
Vous ne trouvez pas le bon produit ?
Essayez notre Outil de sélection de produits.
Certificats d'analyse (COA)
Recherchez un Certificats d'analyse (COA) en saisissant le numéro de lot du produit. Les numéros de lot figurent sur l'étiquette du produit après les mots "Lot" ou "Batch".