HPLC Analysis of Proteins on BIOshell A400 Protein C18
Materiales
columna analítica
Referencia del producto
Descripción
Precios
patrón
Referencia del producto
Descripción
Precios
Trypsin Inhibitor from Glycine max (soybean)
BioUltra, lyophilized powder, ≥95% (Kunitz inhibitor, SDS-PAGE)CONDITIONS
column
BIOShell A400 C18, 10 cm x 2.1 mm I.D., 3.4 μm particles (67463-U)
mobile phase
[A] (water, 0.1% TFA); (acetonitrile, 0.08% TFA), 75:25; [B] (water, 0.1% TFA); (acetonitrile, 0.08% TFA), 25:75
gradient
20 to 100% B in 8 min
flow rate
0.3 mL/min
pressure
2480 psi total pressure (193 bar)
column temp.
30 °C
detector
UV, 215 nm
injection
2 μL
sample
lysozyme, 0.06; haptoglobulin, 0.06; conalbumin, 0.12; tryspin inhibitor, 0.18; glutamate dehydrogenase, (0.18; ß-galactosidase, 0.12; serum albumin chicken, 0.12 g/L in mobile phase A)
Descripción
Nota de análisis
Analysis of proteins of various hydrophobicity on BIOshell A400 Protein C18 to compare the selectivity and retention times.
Información legal
null