Saltar al contenido
Merck

HPLC Analysis of Fat-Soluble Vitamins on Ascentis® Express C18, Methanol Mobile Phase

HPLC Analysis of Fat-Soluble Vitamins on Ascentis® Express C18, Methanol Mobile Phase application for HPLC

Materiales

Estándar

Referencia del producto
Descripción
Precios

Acetato de retinil

analytical standard

Retinol palmitate

analytical standard

Colecalciferol (D3)

analytical standard

Ergocalciferol (D2)

analytical standard

Phylloquinone (K1)

analytical standard

(±)-α-Tocoferol

analytical standard

Vitamin E acetate

analytical standard

δ-Tocopherol

analytical standard

columna analítica

Referencia del producto
Descripción
Precios

Columna para HPLC Ascentis® Express C18, 2,7 μm

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

CONDITIONS

column

Ascentis Express C18, 15 cm x 4.6 mm I.D., 2.7 μm particles (53829-U)

mobile phase

methanol

flow rate

1.5 mL/min

pressure

3249 psi (224 bar)

column temp.

30 °C

detector

UV, 280 nm

injection

5 μL

sample

0.5 mg/mL tocopherol acetate, alpha-tocopherol, delta-tocopherol and K1, 0.05 mg/mL D2, D3 and retinol acetate, 1.5 mg/mL retinol palmitate, all in ethanol.

Descripción

Nota de análisis

Analysis was conducted using standard solution, detection via UV.

Información legal

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany