Saltar al contenido
Merck
  • A study of calcium ion-selective PVC membrane electrode based on neutral carrier N,n,n',n'-tetracyclo-3-oxapentanediamide (correction of oxapetanediamide).

A study of calcium ion-selective PVC membrane electrode based on neutral carrier N,n,n',n'-tetracyclo-3-oxapentanediamide (correction of oxapetanediamide).

Journal of Tongji Medical University = Tong ji yi ke da xue xue bao (1992-01-01)
Y L Ma, X H Rao, S M Zhong, S Ren, T X Yu, Q Zhen
RESUMEN

A calcium ion-selective PVC membrane electrode based on neutral carrier n,n,n',n'-tetracyclohexyl-3-oxapetanediamide, using di-(2-ethylhexyl)phthalate as the plasticizer and potassium tetrakis (4-chlorophenyl) borate as the additive is reported in this paper. The ion selective membrane consists of 1 wt% of the Ca2+ selective ligand, 65 wt% of the plasticizer, 1 wt% of the additive and 33 wt% of poly (vinyl chloride) powder. The electrode has the linear response range of 2.0 x 10(-7)-10(-1) mol/L with the Nernstian slope of 28 mV/decade at 25 degrees C and the detection limit of 2.0 x 10(-8) mol/L. The response time of the calcium ion-selective electrode is as the concentration of calcium ion is rapidly shifted from 10(-5) to 10(-4) mol/L. The potential stability and reproducibility are good. The free calcium in blood serum was determined by the calcium ion-selective electrode with satisfactory results.

MATERIALES
Número de producto
Marca
Descripción del producto

Supelco
Calcium ionophore II, Selectophore, function tested
En este momento no podemos mostrarle ni los precios ni la disponibilidad