UHPLC/MS Analysis of Insulin Analogs on BIOshell A160 Peptide C18

application for UHPLC-MS

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

técnicas

UHPLC-MS: suitable

parámetros analíticos

column: BIOshell A160 Peptide C18, 15 cm x 2.1 mm I.D., 2.0 μm particles (67243-U)
mobile phase: [A] 75:25 (0.1% DFA in water: 0.1% DFA in acetonitrile); [B] 50:50 (0.1% DFA in water: 0.1% DFA in acetonitrile)
gradient: 23% B for 7 min; 23% B to 100% B in 1 min; hold at 100% B for 7 min
flow rate: 0.2 mL/min
pressure: 2500 psi (172 bar)
column temp.: 75 °C
detector: ESI(+), TIC 100-3000 m/z
injection: 0.5 μL
sample: Analog Mixture, 20 μg/mL, 6:3:1 methanol:water:acetic acid containing 0.1% rat plasma

idoneidad

application for UHPLC-MS

aplicaciones

life science and biopharma

Información legal

Analito

Descripción

    Lantus
    Novolog
    Humalog
    Apidra
    Levemir

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documentos section.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico