G005895
LC/MS/MS Analysis of Steroid Hormones in Plasma on Ascentis® Express C18 after Sample Prep using HybridSPE®-Phospholipid
application for HPLC
About This Item
técnicas
HPLC: suitable
parámetros analíticos
SPE tube/cartridge: HybridSPE®-Phospholipid, 96-well plate, 50 mg/well (575656-U)
sample addition: To each well add 100 μL of plasma followed by 300 μL of precipitation solvent (1% formic acid or 0.5% citric acid in acetonitrile). (Agitate via vortex for 4 minute, place on vacuum manifold and apply 10" Hg vacuum for 4 minutes. Collect filtrate and analyze directly.)
column: Ascentis® Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)
gradient: 60% B for 3 min, 60% B to 95% B in 5 min, held at 95% B for 2 min
mobile phase: [A] 5 mM ammonium formate pH 4.0 with formic acid; [B] methanol
flow rate: 0.3 mL/min
column temp.: 50 °C
detector: ESI+, MRM
injection: 2 μL
idoneidad
application for HPLC
aplicaciones
clinical
Nota de análisis
Información legal
Analito
columna analítica
patrón
tubo o placa de SPE
Choose from one of the most recent versions:
Certificados de análisis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico