Saltar al contenido
Merck

LC/MS Analysis of Omeprazole and Metabolites on Ascentis® Express RP Amide (pH 3.4)

LC/MS Analysis of Omeprazole and Metabolites on Ascentis® Express RP Amide (pH 3.4) application for HPLC

Materiales

Analytical column

Referencia del producto
Descripción
Precios

Ascentis® Express RP-Amide, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 5 cm × 2.1 mm

Standard

Referencia del producto
Descripción
Precios

Phenacetin

≥98.0% (HPLC)

Omeprazole sulfone

analytical standard

Omeprazole

analytical standard

CONDITIONS

column

Ascentis Express RP-Amide, 5 cm x 2.1 mm I.D., 2.7 μm particles (53911-U)

mobile phase

[A] 10 mM ammonium formate in water, pH 3.4; [B] 10 mM ammonium formate in methanol, pH 3.4; (65:35, A:B)

flow rate

0.3 mL/min

column temp.

35 °C

detector

ESI(+), m/z 50-400

injection

1 μL

sample

1 mg/L in mobile phase

Descripción

Analysis Note

This application demonstrates the suitability of the Ascentis Express RP Amide for the analysis of omeprazole metabolites.

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany