The recommended alternative product is S2GPU05RE, Stericup Quick Release-GP Sterile Vacuum Filtration System from Millipore. This system includes a 500 mL Steritop bottle top filter unit utilizing a 0.22 µm pore size fast flow PES Express PLUS membrane, along with a 500 mL receiver flask for the sterilization of tissue culture media, protein, and DNA.
Seleccione un Tamaño
| A ustedes/SKU | Disponibilidad | Precio |
|---|
Acerca de este artículo
Quality Segment
material
polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing
sterility
sterile
product line
Stericup®
feature
hydrophilic
packaging
pack of 12 ea
parameter
45 °C max. temp.
description
2.4 mL (hold-up volume)
technique(s)
sterile filtration: suitable
H
263 mm
diam.
12.5 cm
filter capacity
500 mL
filter diam.
73 mm
filtration area
40 cm2
matrix
Express PLUS™
pore size
0.22 μm pore size
fitting
inlet funnel, outlet flask (Receiver)
shipped in
ambient
storage temp.
15-25°C
General description
Application
- Medios de cultivo de tejidos +/- aditivos
- Amortiguadores
- Disoluciones biológicas
Features and Benefits
- El embudo de filtración Quick Release se desconecta de la botella receptora con solo un cuarto de vuelta—reduciendo así la probabilidad de derrame y haciendo que sea más fácil Manipularlo en la mesa de trabajo y debajo de la campana.
- La identificación del producto en negrita en el embudo de filtración Steritop® muestra claramente las características de la membrana, como su composición y el tamaño de los poros, una selección de productos rápida y apropiada para la aplicación.
- El tapón de cierre mediante un clic está bridado y modificado para permitir un fuerte agarre en condiciones secas—y un tope táctil confirma el cierre seguro. Preocúpese menos de la posibilidad de derramar o contaminar el contenido esterilizado.
- La amplia superficie esmerilada de escritura en la botella receptora y el color más claro del tapón facilitan el etiquetado claro para mejorar la legibilidad y reducir los frotis ilegibles.
- La tapa descansa sobre el lateral para evitar el riesgo de contaminación.
- Las botellas receptoras son apilables para ahorrar espacio de almacenamiento.
Legal Information
1 of 1
Este artículo | |||
|---|---|---|---|
| description pore size 0.22 μm, polyethersulfone membrane, filter capacity 500 mL, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 150 mL, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 250 mL, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 1000 mL, pack of 12 ea |
| sterility sterile | sterility - | sterility sterile | sterility sterile |
| filter capacity 500 mL | filter capacity 150 mL | filter capacity 250 mL | filter capacity 1000 mL |
| pore size 0.22 μm pore size | pore size 0.22 μm pore size | pore size 0.22 μm pore size | pore size 0.22 μm pore size |
| material polyethersulfone membrane, polystyrene funnel(s), polystyrene bottle, polystyrene housing | material polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing | material polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing | material polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing |
| packaging pack of 12 ea | packaging pack of 12 ea | packaging pack of 12 ea | packaging pack of 12 ea |
| fitting inlet funnel, outlet flask (Receiver) | fitting inlet funnel, outlet flask (Receiver) | fitting inlet funnel, outlet flask (Receiver) | fitting inlet funnel, outlet flask (Receiver) |
Still not finding the right product?
Explore all of our products under Sistema de filtración al vacío Millipore® Stericup® Quick Release
Elija entre una de las versiones más recientes:
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
-
What is the Corning alternative for the product CLS431097, which is a Corning® bottle-top vacuum filter system with specific features like a polyethersulfone membrane, pore size of 0.22 μm, membrane area of 33.2 cm2, and a filter capacity of 500 mL?
1 answer-
Helpful?
-
-
Can Stericup® bottles be placed inside freezers for long-term storage of media?
1 answer-
You can successfully freeze and store many aqueous solutions (such as culture media) in Stericup® bottles at temperatures up to -20 °C (-4 °F). It is strongly recommended that you run a sample stability trial under your actual storage conditions prior to using Stericup® bottles for frozen storage.
Helpful?
-
-
Can Stericup® and Steritop® filters be reused?
1 answer-
All Stericup® and Steritop® filters were developed for single use, disposable processing.
Helpful?
-
-
What tubing can be used with Stericup® and Steritop® vacuum filters?
What tubing can be used with Stericup® and Steritop® vacuum filters?
1 answer-
Stericup® and Steritop® filters are fitted with a 1/4” tube adapter. We recommend 3/16 in. (4.8 mm) ID silicone tubing to fit this tube adapter: https://www.sigmaaldrich.com/product/mm/xx7100004
Helpful?
-
-
Can we filter oil-type liquids with Stericup® filters?
Can we filter oil-type liquids with Stericup® filters?
1 answer-
Likely not, due to pore size/possible fouling/chemical compatibility. Refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
Helpful?
-
-
Are Stericup® and Steritop® filters GMP-certified?
1 answer-
No, Stericup® and Steritop® devices are not GMP-approved.
Helpful?
-
-
What is the difference between classic Stericup® filters and Stericup® “Quick Release” filters?
1 answer-
Stericup® Quick Release, or “QR”, filters feature a connection between funnel filter and receiver bottle that requires only a quarter-turn release to facilitate funnel removal. These filters come fully engaged – there is no need to tighten before use. The receiver bottle cap has flanges on the top surface to improve grip, as well as a positive lock stop to ensure a leak-proof seal with the receiver bottle. Other improvements include a large, frosted writing surface on the receiver bottle and a lighter cap color to facilitate labelling, more prominent volume markings on the filtration funnel and receiver bottle, and a product label with a peel-off lot record strip on the bag for easy viewing in storage.
Helpful?
-
-
What is the maximum vacuum pressure for filtration?
1 answer-
700 mm Hg (27.5” Hg, 0.9 bar) differential vacuum at 25 °C (77 °F)
Helpful?
-
-
What is the expiration date of Stericup® and Steritop® filters?
1 answer-
Typically 3 years from sterilization.
Helpful?
-
-
What leachables are detected during filtration processes using Stericup® filters?
1 answer-
We have not performed extractables/leachables studies for our Stericup® filters.
Helpful?
-