HPLC Analysis of Spice Cannabinoids JWH-018, JWH-073 and Metabolites on Ascentis® Express C18

application for HPLC

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

HPLC: suitable

Paramètres du test

column: Ascentis Express C18, 10 cm x 3.0 mm I.D., 2.7 μm particles (53814-U)
mobile phase: [A] 0.1% formic acid in water; [B] 0.1% formic acid in water:acetonitrile (10:90)
gradient: 20% B held for 3 min; 20 to 100% B in 0.1 min; 100% B held for 2 min
flow rate: 0.5 mL/min
pressure: 2150 psi (148 bar)
column temp.: 35 °C
detector: UV, 315 nm
injection: 2 μL
sample: 17 μg/mL each in acetonitrile

Adéquation

application for HPLC

Application(s)

forensics and toxicology

Remarque sur l'analyse

Synthetic cannabinoids (e.g. "Spice") are a type of designer drug that provides a cannabis-type high. New synthetic cannabinoids are continually being introduced as suppliers tweak the molecular structures. The ability to rapidly and reliably identify the continually changing population of these compounds is a significant analytical challenge facing forensic chemists. A rapid separation of six of these compounds on an Ascentis Express C18 column is shown here. The highest grade LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    JWH-073 4-Butanoic acid metabolite solution, 100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    JWH-018 5-Pentanoic acid metabolite solution, 100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    JWH-073 solution, 100 μg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®
    JWH-018 solution, 100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Faites votre choix parmi les versions les plus récentes :

Certificats d'analyse (COA)

Lot/Batch Number

Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.

Si vous avez besoin d'assistance, veuillez contacter Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique