Yes, extending the spin times is sufficient. However, the time range would have to be experimentally determined by the user.
Sélectionner une taille de conditionnement
| À vous/Référence | Disponibilité | Prix |
|---|
A propos de cet article
Quality Segment
material
polyethylene cap, polyethylene liner, polypropylene tube, regenerated cellulose membrane, regenerated cellulose support, styrene-butadiene housing
sterility
non-sterile
product line
Amicon®
packaging
pkg of 24 pkg, pkg of 8 pkg, pkg of 96 pkg
manufacturer/tradename
Amicon® Ultra
sustainability
Greener Alternative Product
parameter
4 mL sample volume
technique(s)
protein extraction: suitable, protein purification: suitable
L
124 mm (4.88 in.)
diam.
17.3 mm
filtration area
3 cm2
volume
4 mL
working volume
≤4 mL , 4 mL
pore size
10 kDa MWCO
greener alternative category
shipped in
ambient
General description
Leur configuration verticale et leur surface utile de filtration en font des dispositifs très performants : traitement rapide, très bon taux de récupération (en général plus de 90 % de la solution de départ diluée) et possibilité de concentrer l'échantillon 80 ×. Leur configuration verticale permet en outre de minimiser la polarisation du soluté, et donc, le colmatage de la membrane qui en découle. Enfin, la présence d'une butée physique à l'intérieur du dispositif empêche d'atteindre la siccité tout en prévenant les éventuelles pertes d'échantillon.
Le concentré est récupéré dans le réservoir à échantillon du filtre à l'aide d'une pipette, tandis que l'ultrafiltrat est recueilli dans le tube à centrifuger fourni. Le filtre peut être centrifugé dans un rotor à godets oscillants ou à angle fixe. Avec leur membrane verticale et leur butée fixe, les filtres à centrifuger Amicon® Ultra de 4 ml offrent un taux de récupération maximal pour un temps de centrifugation relativement court.
Application
- Ultrafiltration sur membrane Ultracel® à seuil de coupure de 10 000 Da
- Concentration d'échantillons biologiques contenant des antigènes, des anticorps, des enzymes, des acides nucléiques ou des micro-organismes
- Purification des composants macromoléculaires présents dans des extraits de culture tissulaire ou des lysats cellulaires et élimination des protéines avant chromatographie liquide haute performance
- Concentration de protéines diluées ou prépurifiées à partir d'éluants de colonne
- Dessalage, échange de tampon et dialyse des protéines
- Exclusion stérique et fractionnement des protéines
Features and Benefits
- Membrane Ultracel® en cellulose régénérée à seuil de coupure nominal de 10 kDa offrant une récupération du rétentat supérieure à 90 %
- Réduction de la polarisation de la concentration grâce à la membrane verticale, pour des temps de centrifugation ultrarapides (à peine 5 minutes)
- Membrane thermoscellée minimisant les quantités de matières extractibles en aval
- Tests d'intégrité réalisés sur 100 % des filtres pour garantir la fiabilité des résultats
- Visualisation pratique de l'échantillon grâce au tube translucide et aux graduations de volume
- Pipetage direct de l'échantillon : suppression de l'étape de récupération du concentré
- Facteurs de concentration élevés (80 × à 100 ×)
Legal Information
1 of 1
Cet article | |||
|---|---|---|---|
| description sample volume 4 mL, regenerated cellulose membrane, MWCO 10 kDa | description sample volume 2 mL, regenerated cellulose membrane, MWCO 10 kDa | description sample volume 15 mL, regenerated cellulose membrane, MWCO 10 kDa | description sample volume 0.5 mL, regenerated cellulose membrane, MWCO 10 kDa |
| manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra |
| sterility non-sterile | sterility non-sterile | sterility non-sterile | sterility non-sterile |
| pore size 10 kDa MWCO | pore size 10 kDa MWCO | pore size 10 kDa MWCO | pore size 10 kDa MWCO |
| parameter 4 mL sample volume | parameter 2 mL sample volume | parameter 15 mL sample volume | parameter 0.5 mL sample volume |
| material polyethylene cap, polypropylene tube, regenerated cellulose support, polyethylene liner, styrene-butadiene housing, regenerated cellulose membrane | material polypropylene tube (for collection), regenerated cellulose membrane, styrene-butadiene housing | material polyethylene cap, polyethylene liner, polypropylene tube, regenerated cellulose membrane, regenerated cellulose support, styrene-butadiene housing | material polypropylene cap, polypropylene tube, regenerated cellulose membrane, styrene-butadiene housing |
| working volume ≤4 mL | working volume ≤2 mL | working volume ≤15 mL | working volume ≤500 μL , 0.5 mL |
Still not finding the right product?
Explore all of our products under Filtre à centrifuger Amicon® Ultra, MWCO de 10 kDa
Classe de stockage
10-13 - German Storage Class 10 to 13
Faites votre choix parmi les versions les plus récentes :
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
-
If my centrifuge only goes to 3,052 g, is it okay to centrifuge for a longer period of time to get the same separation? It is a swing-bucket centrifuge using the 10 MWCO filters.
1 answer-
Helpful?
-
-
is-it possible to wash and re-use the Amicon@Ultra-4 ? if yes how many time is_it possible to use them ?
1 answer-
The unit is for one-time use only.
Helpful?
-
-
How should the tubes be sterilized without damaging the membrane?
1 answer-
There are no validated sterilization methods for the Amicon Ultra devices. However, these devices can be "sanitized" with a spin cycle with 70% ethanol, followed by 2 rinsing spins with sterile water. It is recommended to filter the concentrated sample using a sterile 0.2 µm syringe filter (e.g., SLGVR33RS) after processing on the Amicons.
Helpful?
-
-
What speed can this product be centrifuged at ?
1 answer-
For this item when using a swinging-bucket rotor, spin the device at 4,000 × g maximum for approximately 10-40 minutes. When using a fixed-angle rotor, orient the device with the membrane panel facing up. For Amicon® Ultra 3K, 10K, 30K, and 50K devices, spin at 7,500 × g maximum for approximately 10-40 minutes. For Amicon® Ultra 100K devices, spin at 5,000 × g maximum for approximately 10-20 minutes.
Helpful?
-
-
How to use it...how much sample should be loaded and what will be the protocol of centrifugation while using?
1 answer-
The maximum initial sample 4 volume (mL). For more information please see page 36 in this product https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/260/181/br5002en-protein-prep-workflow-ms.pdf
Helpful?
-