Accéder au contenu
Merck
Toutes les photos(1)

Key Documents

KHGLS3THH1

Millipore

Capsule Opticap® XLT multicouche, Durapore® 0,45/0,22 µm

Opticap® XLT 30, cartridge nominal length 30 in. (75 cm), filter diam. 4.2 in. (10.7 cm)

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Code UNSPSC :
23151806

Matériaux

polyester support
polyethylene support
polypropylene filter
polypropylene housing
silicone seal

Niveau de qualité

Conformité réglementaire

FDA 21CFR177-182 (all component materials)

Stérilité

sterile; irradiated

Gamme de produits

Opticap® XLT 30

Caractéristiques

hydrophilic

Fabricant/nom de marque

Opticap®

Paramètres

≤45.0 mL/min air diffusion at 2.75 bar (40 psig) and 23 °C (in water)
1 bar max. inlet pressure (15 psi) at 80 °C
1.05 bar max. differential pressure (15 psid) at 80 °C (Forward:)
2.75 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
3.5 bar max. differential pressure (50 psid) at 25 °C (Reverse; intermittent)
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward:)
5.5 bar max. inlet pressure (80 psi) at 23 °C
80 psig max. inlet pressure

Technique(s)

sterile filtration: suitable

Longueur nominale des cartouches

30 in. (75 cm)

device L

87.1 cm (34.3 in.)

device size

30 in.

Diam. du filtre

4.2 in. (10.7 cm)

filter filtration area

0.1 m2

Diam. du raccord d'entrée

5/8 in.

inlet to outlet width

19.8 cm (7.8 in.)

Diam. du raccord de sortie

5/8 in.

Impuretés

≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL bacterial endotoxins (LAL test, sample aqueous extraction)

Matrice

Multilayer Durapore®

pore size

0.45/0.22 μm pore size

Point de bulle

≥3450 mbar (50 psig), air with water at 23 °C

Raccords

1/4 in. fitting (vent/drain hose barb with double O-ring seal)
(16 mm (5?8 in.) Hose Barb Inlet and Outlet)

Vous recherchez des produits similaires ? Visite Guide de comparaison des produits

Description générale

Device Configuration: Capsule

Conditionnement

Double Easy-Open bag

Notes préparatoires

Sterilization Method:
3 autoclave cycles of 60min at 126°C; not in-line steam sterilizable

This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21CFR210.3(b)(6).

Remarque sur l'analyse

Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2 using ASTM® F838 methodology.
TOC/Conductivity: After sterilization and a controlled water flush, samples exhibited less than 500ppb TOC per USP < 643 > and less than 1.3μS/cm per USP < 645 > at 25°C.
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥ 4.5 L

Autres remarques

Directions for Use:

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

Informations légales

ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
OPTICAP is a registered trademark of Merck KGaA, Darmstadt, Germany

Vous ne trouvez pas le bon produit ?  

Essayez notre Outil de sélection de produits.


Certificats d'analyse (COA)

Recherchez un Certificats d'analyse (COA) en saisissant le numéro de lot du produit. Les numéros de lot figurent sur l'étiquette du produit après les mots "Lot" ou "Batch".

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique