Laissez-nous vous aiderNom du produit
Sucrose (Saccharose) Test, reflectometric, 0.25-2.50 g/L (sucrose), for use with REFLECTOQUANT®
description
quantitative
Quality Segment
analyte chemical class(es)
sugars
reaction suitability
reaction type: enzymatic reaction
measuring range
0.25-2.50 g/L (sucrose)
application(s)
beverage testing
food testing
milk and dairy products
compatibility
RQflex® 20, for use with REFLECTOQUANT®
detection method
reflectometric
storage temp.
2-8°C
General description
Method: reflectometric with test strips and reagent 0.25 - 2.50 g/l Reflectoquant®
The Reflectoquant Sucrose test is suitable for determination of sucrose in beverages, e. g. fruit/vegetable juice and refreshment drinks after appropriate sample pretreatment or dilution. Sucrose (saccharose) is cleaved by sucrose phosphorylase into fructose and glucose-1-phosphate. The latter is converted by phosphoglucomutase into glucose-6-phosphate. This is in turn oxidized by NAD under the catalytic effect of glucose-6-phosphate dehydrogenase to gluconate-6-phosphate. In the presence of diaphorase the NADH formed in the process reduces a tetrazolium salt to a blue formazan that is determined reflectometrically on the RQflex® 20. For the determination of only glucose please refer to our MQuant or Reflectoquant Glucose tests.
Application
Sucrose (Saccharose) Test Strips, reflectometric can be used:
</UL>
In the beverage industry, particularly for determining sucrose levels in soft drinks and fruit juices. This helps manufacturers ensure product consistency and compliance with
</UL>
In the beverage industry, particularly for determining sucrose levels in soft drinks and fruit juices. This helps manufacturers ensure product consistency and compliance with
Analysis Note
function control: entspricht
Assay
- standard 1:
0.26 - 0.34 mg/l
- standard 2: 1.12 - 1.38 mg/l
- standard 3: 1.80 - 2.20 mg/l
Assay
- standard 1:
0.26 - 0.34 mg/l
- standard 2: 1.12 - 1.38 mg/l
- standard 3: 1.80 - 2.20 mg/l
Other Notes
Replaces: 1161410007; 1.16141.0007
Legal Information
REFLECTOQUANT is a registered trademark of Merck KGaA, Darmstadt, Germany
RQFLEX is a registered trademark of Merck KGaA, Darmstadt, Germany
Classe de stockage
10-13 - German Storage Class 10 to 13
Certificats d'analyse (COA)
Recherchez un Certificats d'analyse (COA) en saisissant le numéro de lot du produit. Les numéros de lot figurent sur l'étiquette du produit après les mots "Lot" ou "Batch".
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Numéro d'article de commerce international
| Référence | GTIN |
|---|---|
| 1161410001 | 04022536965710 |