LC/MS/MS Analysis of DL-d-Hydroxyglutaric Acid Enantiomers on Astec® CHIROBIOTIC® R

application for HPLC

Sign Into View Organizational & Contract Pricing

Select a Size


Select a Size

Change View

About This Item


technique(s)

HPLC: suitable

test parameters

column: CHIROBIOTIC R, 25 cm x 4.6 mm I.D., 5 μm particles (13024AST)
mobile phase: (A) ethanol:methanol (3:1, v:v); (B) 0.1% v/v TEA in water, adjusted to pH 4.5 with acetic acid; 75:25 (A:B)
flow rate: 0.4 mL/min
column temp.: 20 °C
detector: MS, ESI(-), 100-1000 m/z
injection: 5 μL
sample: each enantiomer, 100 μg/mL in water

suitability

application for HPLC

application(s)

clinical
forensics and toxicology
life science and biopharma

Analysis Note

Chiral 2-hydroxyglutarates (2-OHG) are important molecular signatures of both healthy and diseased biological cells, their specific biochemical pathways and inborn errors of metabolism. The chiral differentiation and quantification of D-2-OHG (R(-)-2-OHG) and L-2-OHG (S(+)-2-OHG) is key for characterizing neurometabolic disorders like the 2-hydroxyglutaric acidurias that cause neurological impairment early in life. The ability to distinguish and quantify the enantiomers of 2-OHG is therefore important to be able to track its disposition throughout the metabolic pathway of normal vs. mutated cells. We show here a rapid and sensitive method to separate and measure D-2-OHG and L-2-OHG using LC-MS/MS detection. An Astec® CHIROBIOTIC R (ristocetin chiral selector) column was run under polar ionic mobile phase conditions. Using tandem MS eliminated two fundamental analytical challenges with these compounds: the lack of a UV-absorbing chromophore in the 2-OHG molecule, and the abundance of competing low molecular weight acids in urine. The upper panel in the chromatogram shows the separation of D-2-OHG and L-2-OHG obtained during method development. The lower panel shows the pure L-2-hydroxyglutaric acid.
Besides providing the necessary enantioselectivity, Astec® CHIROBIOTIC columns have the advantage of operating in aqueous and polar organic mobile phases that are amenable to polar drugs and metabolites. Their ionic interactions promote analyte ionization which enhances sensitivity in ESI detection. In addition to the CHIROBIOTIC columns, Sigma-Aldrich also provides the high purity mobile phase agents, racemic and pure chiral 2-OHG enantiomers and other compounds in the isocitrate dehydrogenase (IDH) pathway.

Legal Information

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC

Analyte

Description

    L-α-Hydroxyglutaric acid disodium salt, ≥98.0% (GC)
    D-α-Hydroxyglutaric acid disodium salt, ≥95% (GC)

Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Customer Support.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service