The amount of resin in the tube is five grams. The packing is not measured by volume, though the tube itself is a 15 cc tube.
23906
OMI® Purifier Tube
OMI-2 Purifier Tube
Synonym(s):
OMI®-2 INDICATOR TUBE
About This Item
Recommended Products
Looking for similar products? Visit Product Comparison Guide
General description
- A re-useable tube holder that is installed into the gas delivery system. The use of polycarbonate provides see-through capability along with safety.
- A purifier tube containing an indicating resin (changes color from black to brown when exposed to as little as 1 ppm of moisture or oxygen). The use of glass provides see-through capability and prevents diffusion of room contaminants into the gas stream.
Spent purifier tubes are easily replaced. Simply unscrew the end assembly from the tube holder and replace it with a new purifier tube. The design prevents room air from entering the new tube during installation (protective foil covers on each end are only pierced as the end assembly is screwed back onto the tube holder to complete installation).
Note: This item requires a re-useable tube holder (purchased separately).
Footnote
Legal Information
required but not provided
Signal Word
Danger
Hazard Statements
Precautionary Statements
Hazard Classifications
Eye Dam. 1 - Skin Corr. 1B - Water-react 1
Supplementary Hazards
Storage Class Code
4.3 - Hazardous materials which set free flammable gases upon contact with water
WGK
WGK 3
Flash Point(F)
Not applicable
Flash Point(C)
Not applicable
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Customers Also Viewed
Articles
GC-MS method detects nitrosamines in Valsartan tablets, meeting US FDA guidelines for pharmaceutical quality control.
Related Content
Accurately detect and quantify trace nitrosamines (NDMA, NDEA, NEIPA, NDIPA, NDBA, etc.) in pharmaceutical drugs using our complete product portfolio and application guides for LC-MS and GC-MS. Order high-quality reference standards, columns, filters & more.
-
How much purifier material is in the tube by mass and volume?
1 answer-
Helpful?
-
-
How to determine when the OMI® Purifier Tube -23906 is spent and needs to be changed? Also, what are the connections of the OMI® Purifier Tube -23906?
1 answer-
The rapid color change will indicate potential problems in the purification system. If the change is relatively rapid, the bulk purifier is spent, the carrier gas is contaminated, or there is a leak in the system somewhere upstream.
2. The OMI® Purifier Tube can be connected to 1/8" tubing. For 1/4" tubing, use 1/4" to 1/8" Swagelock® reducer (brass - Cat. No 2-1516, or stainless steel - Cat. No 2-1517).
Helpful?
-
-
Could you confirm if product number 23906 OMI® Purifier Tube should have a product number or SKU on the packaging? If not, how to ensure that it is the correct product?
1 answer-
The product number 23906 OMI® Purifier Tube does not have a product number on the aluminum foil bag due to the use of multiple products with this type of bag. To ensure that the correct product is received, customers are advised to open the aluminum bag to find the glass tube labeled OMI-2 23906. This is done to prevent damage during shipment and to avoid any reactions due to exposure to oxygen.
Helpful?
-
-
Hi, I was wondering if there is a maximum flow rate and pressure rating for the OMI® Purifier Tube filters/holders?
1 answer-
The operational line pressure must not exceed 150 psi. Flow should not exceed 300 mL/min for OMI-2.
Helpful?
-
Active Filters
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service