Direkt zum Inhalt
Merck
HomeGas Chromatography (GC) US EPA Method 8015 (modified): GC Analysis of Diesel Range Organics (DRO) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 µm), Fast GC Analysis

US EPA Method 8015 (modified): GC Analysis of Diesel Range Organics (DRO) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 µm), Fast GC Analysis

US EPA Method 8015 (modified): GC Analysis of Diesel Range Organics (DRO) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm), Fast GC Analysis

Conditions

ColumnSLB-5ms, 20 m x 0.18 mm I.D., 0.18 µm (28564-U)
Oven50 °C (2 min), 20 °C/min to 325 °C (1 min)
Carrier Gashelium, 0.9 mL/min, constant
Sampleeach analyte at 50 µg/mL in methylene chloride
Injection1 µL, splitless (splitter open at 0.5 min)
Liner2 mm I.D., FocusLiner™ inlet liner
Inj. Temp.250 °C
DetectorFID, 350 °C

Description

CategoriesAnalytical Chromatography, GC Applications, Hydrocarbons
Featured IndustryEnvironmental
Other NotesComplementary products:
Syringes
Vials
Gas Purification and Gas Management
Inlet Liners, Glass Wool, and Specialized Hand Tools
Septa and Specialized Hand Tools
Column Ferrules, Nuts, and Specialized Hand Tools
Suitabilityapplication for GC

Analyte

  • 5-α-Androstane
  • Decane
  • Docosane
  • Dodecane
  • Eicosane
  • Hexacosane
  • Hexadecane
  • Methylene chloride
  • Octacosane
  • Octadecane
  • o-Terphenyl
  • Tetracosane
  • Tetradecane
Materials