Direkt zum Inhalt
Merck

G005743

HPLC Analysis of Steroids on Ascentis® Express F5

application for HPLC

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

Methode(n)

HPLC: suitable

Testparameter

column: Ascentis Express F5, 15 cm x 4.6 mm I.D., 2.7 μm particles (53591-U)
mobile phase: [A] water; [B] methanol; (50:50, A:B)
flow rate: 0.8 mL/min
pressure: 2393 psi (165 bar)
column temp.: 35 °C
detector: UV, 240 nm
injection: 5 μL
sample: 10 μg/mL of each in 90: 10, water: methanol

Eignung

application for HPLC

Anwendung(en)

clinical

Hinweis zur Analyse

Sample concentrations at 10 ug/mL in 90:10 water:methanol. Acetonitrile does not provide adequate selectivity on this phase. Mass spec was used for peak identification.

Rechtliche Hinweise

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyt

Beschreibung

    Hydrocortisone
    Prednisolone
    Prednisone

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Standard

Produkt-Nr.
Beschreibung
Preisangaben

Choose from one of the most recent versions:

Analysenzertifikate (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Kundensupport

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.