G000507
GC Analysis of Volatiles in Milk (Before and After Exposure to Sunlight) on Supel-Q PLOT after SPME using a 75 μm PDMS/Carboxen Fiber
suitable for GC, application for SPME
About This Item
Recommended Products
technique(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
test parameters
sample/matrix: 3 g of 2% milk + 10 μL internal standard solution (20 μg/mL 4-methyl-2-pentanone) (9 mL GC vial)
SPME fiber: PDMS/Carboxen, 75 μm (57318)
extraction: headspace, 15 min with constant stirring at 45 °C
desorption process: 5 min, 250 °C
column: Supel-Q PLOT, 30 m × 0.32 mm I.D. (24242)
oven: 70 °C (2 min) to 140 °C at 6 °C/min (2 min hold) then to 220 °C at 6 °C/min (5 min hold)
inj. temp.: 250 °C
detector: GC-MS ion trap, m/z = 33-300
carrier gas: helium, 35 cm/sec
injection: splitless, closed 2 min
liner: 0.75 mm liner
sample: Milk
suitability
application for SPME
application(s)
food and beverages
Analysis Note
Legal Information
Analyte
SPME fiber
analytical column
related product
standard
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.