These are the temperature coefficient of variation tables for the five materials:
•Ref. 06949-500ML | Conductivity standard, conductivity 1000 μS/cm at 25 °C
•Ref. 67233-500ML | Conductivity standard, conductivity 10000 μS/cm at 25 °C
•Ref. 56255-500ML | Conductivity standard, conductivity 50000 μS/cm at 25 °C
•Ref. 59777-500ML | Conductivity standard, conductivity 100000 μS/cm at 25 °C
•Ref. 03729-500ML | Conductivity standard, conductivity 200000 μS/cm at 25 °C
Select a Size
| Size/SKU | Availability | Price |
|---|---|---|
500 mL | Please contact Customer Service for Availability | CZK 5,710.00 |
About This Item
grade
analytical standard
Quality Segment
shelf life
limited shelf life, expiry date on the label
concentration
2.5-10% KCl (T)
conductivity
49500-50500 μS/cm at 25 °C
density
1.014-1.063 g/cm3 at 20 °C
application(s)
cleaning products
cosmetics
environmental
flavors and fragrances
food and beverages
personal care
format
single component solution
Application
1 of 1
This Item | |||
|---|---|---|---|
| format single component solution | format single component solution | format single component solution | format single component solution |
| grade analytical standard | grade analytical standard | grade analytical standard | grade analytical standard |
| application(s) cleaning products | application(s) environmental | application(s) environmental | application(s) cleaning products |
| concentration 2.5-10% KCl (T) | concentration 10-25% KCl (T) | concentration ≤2.5% KCl (T) | concentration ≤2.5% KCl (T) |
| Quality Level 100 | Quality Level 100 | Quality Level 100 | Quality Level 100 |
| density 1.014-1.063 g/cm3 at 20 °C | density 1.063-1.135 g/cm3 at 20 °C | density 0.995-1.014 g/cm3 at 20 °C | density 0.995-1.014 g/cm3 at 20 °C |
Still not finding the right product?
Explore all of our products under Conductivity Standard
signalword
Warning
hcodes
pcodes
Hazard Classifications
Carc. 2 - STOT RE 2
Storage Class
10 - Combustible liquids
wgk
WGK 2
flash_point_f
Not applicable
flash_point_c
Not applicable
Regulatory Listings
Regulatory Listings are mainly provided for chemical products. Only limited information can be provided here for non-chemical products. No entry means none of the components are listed. It is the user’s obligation to ensure the safe and legal use of the product.
67-66-3
Choose from one of the most recent versions:
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
-
Do these products have conductivity correction tables at different temperatures?
1 answer-
Helpful?
-
-
Hi, I wonder if u maybe have some kind of table of conductivity corrections at various temperatures? Thanks in advance!
1 answer-
Unfortunately, there is no table for conductivity at different temperatures available. The only value of conductivity shared is on the lot specific Certificate of Analysis where the Conductivity range is determined at 25 C between 49500 - 50500 µS/cm. The exact value will be listed on the Certificate of Analysis.
Helpful?
-