The reagents used with the MAK311-1KT Peroxide Assay Kit create an acidic environment in the well plate, and Phenol Red should not have measurable absorbance at the kit's wavelength of 585 nm.
As an extra precaution, it is recommended to dilute the H2O2 standard in the same media that is used on the cells to account for any background or chemical interactions from the media. Furthermore, it is best for samples to contain less than 1 mM pyruvate.
MAK311
Peroxide Assay Kit
sufficient for 250 colorimetric tests
Synonym(s):
Hydrogen Peroxide Test Kit
About This Item
Recommended Products
detection method
colorimetric
relevant disease(s)
cardiovascular diseases; aging/geriatric diseases; neonatal diseases; pediatric diseases; orthopedic diseases; neurological disorders; pulmonary disorders
storage temp.
2-8°C
Related Categories
General description
Application
Features and Benefits
Suitability
Principle
Signal Word
Warning
Hazard Statements
Precautionary Statements
Hazard Classifications
Eye Irrit. 2 - Met. Corr. 1 - Skin Irrit. 2
Storage Class Code
8B - Non-combustible corrosive hazardous materials
Flash Point(F)
Not applicable
Flash Point(C)
Not applicable
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Don't see the Right Version?
If you require a particular version, you can look up a specific certificate by the Lot or Batch number.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Customers Also Viewed
-
Is it possible to use the MAK311-1KT Peroxide Assay Kit to measure peroxide in lysed cells that are grown in Minimum Essential Medium Eagle containing Phenol Red at a concentration of 0.011 g/liter?
1 answer-
Helpful?
-
-
Is 1% TWEEN in the sample compatible with the MAK311 peroxide assay kit?
1 answer-
The tween in samples should remain below 0.25%. If the sample contains 1% tween it is recommend diluting 4x to 0.25%, assuming the sample has enough peroxide to still be in the detectable range. The kit has a detection range of 0.2–30 μM (7–1,020 ng/mL) H2O2 in a 96 well format.
Helpful?
-