HPLC Analysis of Alfuzosin and Related Substances on Chromolith® HighResolution RP-18 endcapped

application for HPLC

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

HPLC: suitable

Parametri analitici

sample preparation: Dissolve 40.0 mg of the sample in 100 mL of mobile phase. 0.4 mg/mL (400 ppm) of USP Alfuzosin System Suitability Mixture RS in Mobile phase
column: Chromolith® HighResolution RP-18 endcapped, 150 x 4.6 mm (1.52023.0001)
mobile phase: [A] 5 ml of perchloric acid in 900 mL Milli-Q water. Adjust pH to 3.5 with 2M NaOH. Diluter to 1000 mL with water; [B] acetonitrile; [C] THF; (80:20:1, A:B:C)
flow rate: 1 mL/min
pressure: 994 psi (69 Bar)
column temp.: 25 °C
detector: 254 nm cell: 10 μL
injection: 10 μL
sample: 0.40 μg/mL of Alfuzosin Hydrochloride in Mobile phase from the Sample solution

Compatibilità

application for HPLC

applicazioni

pharmaceutical (small molecule)

Note legali

Chromolith is a registered trademark of Merck KGaA, Darmstadt, Germany

Analita

Descrizione

    Deacylated Alfuzosin
    Alfuzosin
    Furamide analog

Choose from one of the most recent versions:

Certificati d'analisi (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.