G005908
HPLC Analysis of Aflatoxins on Ascentis® Express C18
application for HPLC
About This Item
tecniche
HPLC: suitable
Parametri analitici
column: Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)
mobile phase: [A] water; [B] methanol; [C] acetonitrile; (720:120:120, A:B:C) containing 0.780 g/L potassium bromide, 230 μL/L nitric acid
flow rate: 0.400 mL/min
pressure: 4568 psi (315 bar)
column temp.: 35 °C
detector: FLD excitation = 360 nm, emission = 440 nm
injection: 20 μL
sample: 200 μL spiked 84:16 acetonitrile:di water in 880 μL di water, Spike concentrations: B1 = 16 ppb; B2 = 4 ppb; G1 = 16 ppb; B2 = 4 ppb
Compatibilità
application for HPLC
applicazioni
food and beverages
Risultati analitici
Note legali
Analita
Colonna analitica
Standard
Choose from one of the most recent versions:
Certificati d'analisi (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.