GC Analysis of Parabens on SLB®-5ms after SPME using 50/30 μm DVB/Carboxen/PDMS Fiber

suitable for GC, application for SPME

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

Parametri analitici

sample/matrix: parabens, each at 200 ppb in 3 mL water + 25% sodium chloride in a 4 mL vial
SPME fiber: metal fiber assembly coated with 50/30 μm DVB/Carboxen/PDMS (57912-U)
extraction: immersion with stirring, 25 °C (15 min)
desorption process: 260 °C, 2 min
column: SLB-5ms, 20 m × 0.18 mm I.D., 0.36 μm (28576-U)
oven: 60 °C (2 min), 15 °C/min to 300 °C (5 min)
MSD interface: 275 °C
scan range: m/z 40-450
carrier gas: helium, 0.7 mL/min constant
liner: 0.75 mm I.D. SPME

Compatibilità

application for SPME

applicazioni

chemicals and industrial polymers
cleaning products
cosmetics
food and beverages
personal care
pharmaceutical (small molecule)

Note legali

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany

Analita

Descrizione

    Methyl 4-hydroxybenzoate
    Ethyl 4-hydroxybenzoate
    Propyl 4-hydroxybenzoate
    Butyl 4-hydroxybenzoate

Choose from one of the most recent versions:

Certificati d'analisi (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.