Passa al contenuto
Merck
Tutte le immagini(1)

Documenti fondamentali

G003763

HPLC Analysis of Cortocisteroids in Urine on Discovery® HS F5 after SPE using Discovery® DSC-CN

application for SPE, application for HPLC

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

HPLC: suitable
solid phase extraction (SPE): suitable

Parametri analitici

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: human urine spiked with corticosteroids at 0.5 and 1.0 μg/mL and diluted in DI water (1:1, v/v)
SPE well plate: Discovery DSC-CN 96-well SPE, 100 mg/well (575636-U)
condition: 1 mL methanol; 1 mL DI water.
sample addition: 1 mL spiked urine
washing: 1 mL 20% methanol
elution: 1 mL methanol
eluate post-treatment: Evaporate eluate with nitrogen at 30 °C. Reconstitute in 200 μL mobile phase.
column: Discovery HS F5, 5 cm x 4.6 mm I.D., 3 μm particles (567504-U)
mobile phase: [A] methanol; [B] DI water; (40:60, A:B)
flow rate: 1.5 mL/min
column temp.: 35 °C
detector: UV, 240 nm
injection: 5 μL

Compatibilità

application for HPLC
application for SPE

applicazioni

clinical

Note legali

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documenti section.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.