G003595
GC Analysis of Agricultural Pesticides (Standard) on SLB®-5ms
suitable for GC
About This Item
tecniche
gas chromatography (GC): suitable
Parametri analitici
column: SLB-5ms, 30 m x 0.25 mm I.D, 0.25 μm (28471-U)
oven: 100 °C (1 min), 10 °C/min to 300 °C (5 min)
inj. temp.: 250 °C
MSD interface: 300 °C
scan range: selected ion monitoring (SIM), 7 monitoring groups used
carrier gas: helium, 1 mL/min constant
injection: 1.0 μL pulsed (20 psi until 0.20 min), splitless (1.0 min)
liner: 4 mm I.D., single taper
sample: 29-component pesticide mix plus 1 internal standard, each at 100 ppb in toluene
applicazioni
agriculture
food and beverages
Risultati analitici
standard solution was made in toluene to increase it stability and enhance response of some of the pesticides. The extra peaks in the first portion of the run are due to interferences from the solvent.
Note legali
Analita
Colonna analitica
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.